Wavelength Range | 380 – 780 nm |
---|---|
Luminance Range | 0.002 fL to 6000 fL 0.002 fL @ Signal to Noise Ratio > 10, 20 seconds exposure 0.01 fL @ Signal to Noise Ratio > 10, 0.5 second exposure 0.01 fL @ Signal to Noise Ratio > 100, 20 seconds exposure 6000 fl @ Signal to Noise Ratio > 8000, 0.001 second exposure |
Luminance Accuracy | ± 2 % @ 0.3 fL, 0.5 second exposure |
Luminance Repeatability | 0.2 % @ 0.3 fL, 0.5 second exposure 1.5 % @ 0.03 fL, 0.5 second exposure 1.5 % @ 0.003 fL, 20 seconds exposure |
Chromaticity Accuracy | ± 0.0015 x, y @ 0.5 fL, 0.5 second exposure ± 0.0015 x, y @ 0.1 fL, 20 seconds exposure |
Chromaticity Repeatability | 0.0005 x, y @ 0.5 fL, 0.5 second exposure 0.0005 x, y @ 0.1 fL, 20 seconds exposure |
Polarization error: | ≤ 0.1 % |
Synchronization frequencies: | 10-500 Hz (Custom Synchronization from 10 Hz to 10 KHz available) |
Exposure Time Range: | 1 ms to 20 seconds |
Power Requirements | 5V, 120 mA (600 mW) via USB 2.0 |
Interface | USB 2.0, Ethernet, RS-232 |
Weight / Mass | 1 pound (0.45 kg) 1.2 pounds (0.54 kg) with viewing system VO-201 |